Info & pricing
570-538-7200
tpsinfo@tps.spx.com
|
|
1700° Box Furnaces, Temperature Range 500°C
to 1700°C
| Model |
Independent
Controller |
Electrical
Volts,
Hz, 1Ø |
Watts |
Interior
Dimensions
W X D X H in" (mm) |
Exterior
Dimensions
W X D X H in" (mm) |
Ship
Weight
lbs (kg) |
| BF51314C |
CC59246PCOMC |
208/240V,
50/60Hz |
3500 |
5"(127)x6"(152)x5"(127) |
16"(406)x16"(406)x14"(356) |
85(39) |
| BF51524C |
CC59256PCOMC |
208/240V,
50/60Hz |
3500 |
8.5"(216)x10"(254)x6.5"(165) |
19.5"(495)x19.5"(495)x15.75"(400) |
115(53) |
Controllers
1700°C Controller, Programmable, With
Communications
Lindberg/Blue M 1700°C Programmable Controllers provide multiple
programmable and multiple segments for ramp (up and down) and dwell
(timed holds) temperature control. The controller visually displays
ramp rate, dwell time, program segment and percent power output.
A holdback feature allows the operator to set a "process vs.
setpoint" temperature value which, when exceeded, holds the
program to allow the process to catch up.
The controller includes a selectable self-tuning feature which
automatically sets the best PID settings (proportional, integral,
derivative) for the thermal process. LED display indicates actual
temperature and setpoint temperature. High limit overtemperature
protection is standard. The control console includes a circuit
breaker, power module, transformer and cooling fans.
Programmable Control With Communications
- Available on “COM” Models
- Controllers include RS485
data port (communications card and port) for connection to remote
computer, allowing modification, interrogation and data transfer
of all instrument control and configuration parameter. Up to
30 units can be connected to one PC. Software is not included,
but is available as an option.
- Includes installed communications
card and port for user connection
Option B Overtemperature Control (OTC)
Adjustable digital overtemperature control, factory installed on
selected control consoles with "B" designation; see
chart. Protects furnace and load in the event of primary control
circuit failure. Overrides main controller, assumes temperature
control and shuts off power to furnace if high limit is reached.
Manual re-set required for safety. Operates via signal from independent
thermocouple.
| Model |
CC59246PCOMC |
CC59246PBCOMC |
CC59256PCOMC |
CC59256PBCOMC |
| Digital |
|
|
|
|
| With Programmer |
|
|
|
|
| With Overtemp
Control |
|
|
|
|
| Electrical
Volts, Hz |
208/240V, 50/60Hz |
208/240V, 50/60Hz |
208/240V, 50/60Hz |
208/240V, 50/60Hz |
Exterior
Dimensions
WxDxH in" (mm) |
21"(533)x15"(381)
x10"(254) |
21"(533)x15"(381)
x10"(254) |
21"(533)x15"(381)
x10"(254) |
21"(533)x15"(381)
x10"(254) |
Ship Weight
lbs (kg) |
115(53) |
115(53) |
130(59) |
130(59) |
Notes:
- Maximum weight capacity is 0.5 pounds per square inch of interior
floor surface.
Actual performance may vary depending on load, chamber size, sample
placement, ambient temperature and environmental conditions.
Heat-Up/Cool-Down Rates
|